![Rohde & Schwarz 1321.1211.02 User Manual Download Page 70](http://html1.mh-extra.com/html/rohde-and-schwarz/1321-1211-02/1321-1211-02_user-manual_1477772070.webp)
Basic Operation
R&S
®
Cable Rider ZPH
63
User Manual 1321.0950.02 ─ 07
The connection is now created and configured.
7. Select the new connection labeled R&S Cable Rider ZPH.
8. Select the "Connect" button to establish the connection.
Connecting the R&S Cable Rider ZPH in an existing LAN
You can either get the R&S Cable Rider ZPH IP address automatically from the DHCP
server or manually assign a fixed address.
With manual allocation, a fixed IP address and subnet mask must be assigned to the
R&S Cable Rider ZPH as described in
Chapter 4.3.1, "LAN Connection"
,
on page 60.
After setting up the R&S Cable Rider ZPH, configure the R&S InstrumentView software
with the assigned IP address as described in
"Configuring the R&S InstrumentView
software package"
on page 62.
Free IP address
Contact your IT system administrator to get a free IP address.
In networks with a DHCP server, DHCP permits automatic allocation of the network
configuration to the R&S Cable Rider ZPH connected via LAN cable. For this purpose,
DHCP has to be active on the R&S Cable Rider ZPH.
DHCP is off by default. Turn it on like this:
1. In the "Instrument Setup" dialog box, select the "DHCP" item.
2. Select "DHCP" to "On" to activate DHCP.
Connecting the R&S Cable Rider ZPH to a PC