Remote Commands
R&S
®
Cable Rider ZPH
468
User Manual 1321.0950.02 ─ 07
List of commands
CALCulate<1...2>:FEED
.................................................................................................468
INSTrument[:SELect]
......................................................................................................468
INSTrument:NSELect
..................................................................................................... 469
CALCulate<1...2>:FEED
<Measurement>
This command switches the measurement mode.
The numeric suffix at CALCulate is irrelevant for this command
Suffix:
<1...2>
.
1...2
Parameters:
<Measurement>
String containing the measurement mode (analog modulation or
digital modulation):
Example:
CALC:FEED 'XTIM.AM'
Switches the analog measurement mode to AM Time.
CALC:FEED 'XTIM.FM'
Switches the analog measurement mode to FM Time.
CALC:FEED 'XTIM.FMSummary'
Switches the analog measurement mode to FM Summary.
CALC:FEED 'XTIM:DDEM:ERR:MPH'
Switches the digital measurement mode to error result display.
CALC:FEED 'XTIM:DDEM:EYE'
Switches the digital measurement mode to eye diagram.
CALC:FEED 'XTIM:DDEM:MEAS'
Selects the modulation depth (ASK) or modulation deviation
(FSK) result display.
CALC:FEED 'XTIM:DDEM:SYMB'
Switches the digital measurement mode to symbol stream result
display.
INSTrument[:SELect]
<OperatingMode>
This command selects the operating mode.
Parameters:
<OperatingMode>
SANalyzer | PM | ADEModulation | DDEMod | RECeiver |
MAPS | ACT
SANalyzer
spectrum analyzer
PM
power meter
ADEModulation
analog modulation
DDEMod
digital modulation
Remote Control - Commands