Getting Started
R&S
®
Cable Rider ZPH
29
User Manual 1321.0950.02 ─ 07
RF out / Reflection
For model .02, the RF out connector provides a signal source output power at -10.00
dBm nominal.
For model .12, the RF out connector provides the following ways to generate a signal
source output power at -10.00 dBm nominal.
●
Tacking generator output
●
Continuous signal source output
●
Coupled continuous signal source output
For model .12, the RF out connector can also be configured as an internal DC bias
port. When the RF out is configured as a BIAS port, the RF out connector provides
direct current output. Active equipment under test can be supplied with power via the
bias port.
Risk of electrical shock
To avoid electrical shock, the DC input voltage must never exceed the value specified
on the housing.
Risk of instrument damage
●
Make sure not to overload the RF input and keep within the maximum allowed sig-
nal levels. Refer to the datasheet for the maximum allowed signal levels.
●
To avoid damage to the coupling capacitor, input attenuator or the mixer, the DC
input voltage of 50 V must never be exceeded.
3.2.3
Left View
1 = DC input
2 = Kensington lock slot
Instrument Tour