Basic Operation
R&S
®
Cable Rider ZPH
38
User Manual 1321.0950.02 ─ 07
"Config Overview" Block
Corresponding Dialog Box
Description
Select "Cable" to configure the
cable model used in the consider-
ation of calculation during mea-
surement.
If "User Mode" is set "No", the
preloaded model values are dis-
played for the field as read-only
parameters for "Frequency",
"Veocity Factor" and "Loss".
If "User Model" is set "Yes", the
"Cable Model" is switched to
"UserCableModel" and the last
user-defined values are re-loaded
for the field "Frequency", "Veocity
Factor" and "Loss". These fields
are available for user-specific
cable model setting.
See
Chapter 7.2.1, "Selecting the
Cable Mode"
,
on page 143.
Select "Distance" to configure the
cable length and unit used to
determine the cable length.
See
Chapter 7.2.2.2, "Setting the
Start and Stop Distance for DTF
Measurements"
,
on page 148.
Select "Amplitude" to configure
the parameters related to the ver-
tical axis of the trace window. You
can also select the "Cuurent
Trace" to display in the "Split
Screen" menu item.
The "TG Power" is only available
for model .12.
See
Chapter 7.2.3, "Configuring
the Vertical Axis"
,
on page 150
and
Split screen.
Select "Analysis" to determine the
type of trace mode (Average,
Clear/Write) and the number of
measurement points used in the
measurement. If "Trace Mode" is
configured as "Average", the field
"Avg Count" is available for con-
figuration.
See
Chapter 7.3, "Analyzing Mea-
surement Results"
,
on page 155
and
Chapter 7.2.6.2, "Changing
the Number of Measurement
Points"
,
on page 154.
Screen Layout and Elements