Remote Commands
R&S
®
Cable Rider ZPH
409
User Manual 1321.0950.02 ─ 07
If you are using the average, max hold or min hold trace mode, you can set the number
of measurements with
[SENSe:]SWEep:COUNt<SweepCount>
. Note that synchroni-
zation to the end of the average count is possible only in single sweep mode.
The numeric suffix at DISPlay is irrelevant for this command.
The numeric suffix at TRACe selects the number of the trace.
Parameters:
<DisplayMode>
WRITe | AVERage | MINHold | MAXHold | VIEW | FREeze |
INFinite
*RST:
WRITe
You can turn off the trace with
DISPlay<1...2>[:WINDow]:TRACe<1...2>[:STATe]
.
For more information, see chapter "Trace Mode".
Example:
SWE:CONT OFF
SWE:COUN 16
Turn on single sweep mode and sets the number of measure-
ments to 16.
DISP:TRAC:MODE MAXH
Activates MAXHold mode for the trace.
INIT;*WAI
Starts the measurement and waits for the end of the 16 sweeps.
FORMat:BORDer
<TransferOrder>
This command selects the format of binary data.
Parameters:
<TransferOrder>
NORMal | SWAPped
NORMal
The most significant byte is transferred first big endian).
SWAPped
The least significant byte is transferred first (little endian).
*RST:
SWAPped
Example:
FORM:BORD NORM
Changes the byte order to normal mode.
FORMat[:DATA]
<Format>[, <>]
This command selects the data format that is used for transmission of trace data from
the R&S Cable Rider ZPH to the controlling computer.
Note that the command has no effect for data that you send to the R&S Cable Rider
ZPH. The R&S Cable Rider ZPH automatically recognizes the data it receives, regard-
less of the format.
Parameters:
<Format>
ASCii | REAL
Remote Control - Commands