Remote Commands
R&S
®
Cable Rider ZPH
423
User Manual 1321.0950.02 ─ 07
This automatic routine makes sure that the signal power level does not overload the
R&S Cable Rider ZPH or limits the dynamic range by too small an S/N ratio.
To determine the best reference level, the R&S Cable Rider ZPH aborts current mea-
surements and performs a series of test sweeps. After it has finished the test, it contin-
ues with the actual measurement.
This command is an event and therefore has no *RST value and no query.
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
MARKer selects the marker.
Example:
CALC:MARK:FUNC:LEV:ONCE
Initiates an automatic level adjust routine.
Usage:
Event
CALCulate<1...2>:MARKer<1...6>:FUNCtion:NDBDown
<MarkerDistance>
This command defines the distance of the n dB down markers to the reference marker.
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
MARKer selects the marker.
Parameters:
<MarkerDistance>
Distance of the temporary markers to the reference marker in
dB.
*RST:
3 dB
Default unit: dB
Example:
See
CALCulate<1...2>:MARKer<1...6>:FUNCtion:
NDBDown:STATe
on page 368
CALCulate<1...2>:MARKer<1...6>:FUNCtion:NDBDown:FREQuency?
This command queries the horizontal position of the n dB down markers.
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
MARKer selects the marker.
Parameters:
<Return values>
<frequency1>, <frequency2>
<frequency1>
Absolute frequency of the n dB marker to the left of the refer-
ence marker in Hz.
<frequency2>
Absolute frequency of the n dB marker to the right of the refer-
ence marker in Hz.
Example:
See
CALCulate<1...2>:MARKer<1...6>:FUNCtion:
NDBDown:STATe
on page 368
Usage:
Query only