![Rohde & Schwarz 1321.1211.02 User Manual Download Page 442](http://html1.mh-extra.com/html/rohde-and-schwarz/1321-1211-02/1321-1211-02_user-manual_1477772442.webp)
Remote Commands
R&S
®
Cable Rider ZPH
435
User Manual 1321.0950.02 ─ 07
Parameters:
<ChannelTable>
String containing the name of the channel table.
*RST:
' '
Example:
See
[SENSe:]CHANnel:TABLe:SELect<1...4>
on page 433.
[SENSe:]CHANnel:TABLe:SET
<Table>
This command selects the link direction for measurements with channel tables.
Parameters:
<Table>
UP | DOWN
UP
Selects the uplink.
DOWN
Selects the downlink.
*RST:
UP
Example:
See
[SENSe:]CHANnel:TABLe:SELect<1...4>
on page 433.
Power Measurements
The following commands configure power measurements. To perform the actual mea-
surement, use the commands described in section
Chapter 14.6.3.5, "Performing and
Triggering Measurements"
,
on page 402.
List of commands
CALCulate<1...2>:MARKer<1...6>:FUNCtion:POWer:SELect
............................................. 435
CALCulate<1...2>:MARKer<1...6>:FUNCtion:POWer[:STATe]
.............................................436
CALCulate<1...2>:MARKer<1...6>:FUNCtion:POWer:PRESet
............................................ 436
CALCulate<1...2>:MARKer<1...6>:FUNCtion:POWer:PRESet:CHECk?
...............................437
CALCulate<1...2>:MARKer<1...6>:FUNCtion:LEVel:ONCE
.................................................437
CALCulate<1...2>:MARKer<1...6>:FUNCtion:POWer:RESult?
............................................437
CALCulate<1...2>:MARKer<1...6>:FUNCtion:POWer:SELect
<MeasType>
This command selects a power measurement and turns the measurement on.
The numeric suffix at CALculate and MARKer are irrelevant for this command.
Parameters:
<MeasType>
MCACpower | ACPower | CPOWer | OBANdwidth | OBWidth |
TDMA
MCACpower | ACPower
Adjacent Channel Leakage Ratio measurement (ACLR) aka
adjacent channel power or multi carrier adjacent channel.
CPOWer
Chanel power measurement.
Remote Control - Commands