Remote Commands
R&S
®
Cable Rider ZPH
402
User Manual 1321.0950.02 ─ 07
14.6.3.5
Performing and Triggering Measurements
The following commands control the actual measurement process, including trigger
functionality.
●
Performing the Measurement
................................................................................402
●
Triggering Measurements
..................................................................................... 404
Performing the Measurement
The following commands initialize a measurement and setup the sweep.
List of commands
*WAI
............................................................................................................................. 402
ABORt
.......................................................................................................................... 402
INITiate[:IMMediate]
....................................................................................................... 402
[SENSe:]SWEep:COUNt
.................................................................................................403
INITiate:CONTinuous
......................................................................................................403
[SENSe:]SWEep:POINts
.................................................................................................403
[SENSe:]SWEep:TIME
................................................................................................... 404
[SENSe:]SWEep:TIME:AUTO
......................................................................................... 404
*WAI
Wait to continue
Prevents servicing of the subsequent commands until all preceding commands have
been executed and all signals have settled (see also command synchronization and
*OPC
).
Usage:
Event
ABORt
This command aborts the current measurement and resets the trigger system.
This command is an event and therefore has no query and no *RST value.
Example:
ABOR;
INIT:IMM
Aborts a measurement and starts a new one.
Usage:
Event
INITiate[:IMMediate]
This command initiates a new measurement sequence.
With sweep count > 0 or average count > 0, this means a restart of the indicated num-
ber of measurements. With trace functions MAXHold, MINHold and AVERage, the pre-
vious results are reset on restarting the measurement.