Remote Commands
R&S
®
Cable Rider ZPH
364
User Manual 1321.0950.02 ─ 07
CALCulate<1...2>:MARKer<1...6>:FUNCtion:DEModulation[:STATe]
................................... 365
CALCulate<1...2>:MARKer<1...6>:FUNCtion:DEModulation:HOLD
..................................... 366
CALCulate<1...2>:MARKer<1...6>:FUNCtion:DEModulation:SELect
....................................366
CALCulate<1...2>:MARKer<1...6>:FUNCtion:LEVel:ONCE
.................................................366
CALCulate<1...2>:MARKer<1...6>:FUNCtion:NDBDown
.................................................... 367
CALCulate<1...2>:MARKer<1...6>:FUNCtion:NDBDown:FREQuency?
................................367
CALCulate<1...2>:MARKer<1...6>:FUNCtion:NDBDown:RESult?
....................................... 368
CALCulate<1...2>:MARKer<1...6>:FUNCtion:NDBDown:STATe
.......................................... 368
CALCulate<1...2>:MARKer<1...6>:FUNCtion:NOISe[:STATe]
..............................................368
CALCulate<1...2>:MARKer<1...6>:FUNCtion:NOISe:RESult?
.............................................369
CALCulate<1...2>:MARKer<1...6>:FUNCtion:REFerence
...................................................369
CALCulate<1...2>:MARKer<1...6>:COUNt:FREQuency?
This command performs a frequency measurement at the marker position and returns
the result.
To get a valid result, you have to perform a complete sweep with synchronization to the
sweep end to make sure that the R&S Cable Rider ZPH actually reaches the frequency
you want to measure. This is only possible in single sweep mode.
Before you can use the command, you have to turn on the frequency counter with
CALCulate<1...2>:MARKer<1...6>:COUNt[:STATe]
on page 364.
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
MARKer selects the marker.
Example:
See
CALCulate<1...2>:MARKer<1...6>:COUNt[:STATe]
on page 364.
Usage:
Query only
CALCulate<1...2>:MARKer<1...6>:COUNt[:STATe]
<State>
This command turns the frequency counter at the marker position on and off.
You can read out the result with
CALCulate<1...2>:MARKer<1...6>:COUNt:
FREQuency?
on page 364.
Frequency counting is possible only for one marker at a time. If it is activated for
another marker, it is automatically deactivated for the previous marker.
To get a valid result, you have to perform a complete sweep with synchronization to the
sweep end to make sure that the R&S Cable Rider ZPH actually reaches the frequency
you want to measure. This is only possible in single sweep mode.
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
MARKer selects the marker.
Parameters:
<State>
ON | OFF
*RST:
OFF
Remote Control - Commands