Remote Commands
R&S
®
Cable Rider ZPH
478
User Manual 1321.0950.02 ─ 07
INITiate[:IMMediate]
This command initiates a new measurement sequence.
With sweep count > 0 or average count > 0, this means a restart of the indicated num-
ber of measurements. With trace functions MAXHold, MINHold and AVERage, the pre-
vious results are reset on restarting the measurement.
In single sweep mode, synchronization to the end of the indicated number of measure-
ments can be achieved with the command *OPC, *OPC? or *WAI. In continuous–
sweep mode, synchronization to the sweep end is not possible since the overall mea-
surement never ends.
This command is an event and therefore has no query and no *RST value.
Example:
INIT:CONT OFF
DISP:WIND:TRAC:MODE AVER
Turns on single sweep mode and trace averaging.
INIT;*WAI
Starts the measurement and waits for the end of the sweep.
Usage:
Event
INITiate:CONTinuous
<State>
This command selects the sweep mode.
Parameters:
<State>
ON | OFF
ON
Selects continuous sweeps.
OFF
Selects single sweep.
*RST:
ON
Example:
INIT:CONT OFF
Turns on single sweep mode.
14.6.5.4
Configuring the Trigger
The following chapter describes command necessary to configure the trigger signal for
the digital modulation parameters.
List of commands
TRIGger[:SEQuence]:SLOPe
.......................................................................................... 478
TRIGger[:SEQuence]:LEVel:IQPower
...............................................................................479
TRIGger[:SEQuence]:SOURce
........................................................................................479
TRIGger[:SEQuence]:SLOPe
<TriggerSlope>
This command selects the slope of the trigger signal.
Remote Control - Commands