Remote Commands
R&S
®
Cable Rider ZPH
398
User Manual 1321.0950.02 ─ 07
Example:
SOUR:TG:ATT 40
Sets the attenuation to 40 dB and therefore an output level of –
40 dBm
SOURce:TG:AUTO
<TgSourceMode>
This command turns the generation of a CW signal in zero span on and off.
In case of zero span, you can define a fixed frequency for the generated signal
SOURce:TG:FREQuency
and turn the tracking generator on and off (
SOURce:TG[:
STATe]
).
Parameters:
<TgSourceMode>
ON | OFF
*RST:
OFF
Example:
SOUR:TG:AUTO ON
Selects zero span measurement mode.
SOURce:TG:FREQuency
<TgSourceFrequency>
This command defines the frequency of the generated signal.
Available when zero span has been turned on, for example with
SOURce:TG:AUTO
.
Parameters:
TgSrcFreq
Frequency of the CW signal.
*RST:
OFF
Default unit: Hz
Example:
SOUR:TG:AUTO ON
SOUR:TG:FREQ 500MHZ
SOUR:TG:POW -10
Generates a CW signal with 500 MHz and a power of -10 dBm in zero span.
SOURce:TG:FREQuency:AUTO
<TgAutoFrequency>
This command coupled the frequency of the CW signal as the center frequency of the
current measurement.
Parameters:
<TgAutoFrequency> *RST:
OFF
Default unit: Hz
Example:
SOUR:TG:AUTO ON
SOUR:TG:FREQ:AUTO ON
Generates a CW signal with frequency coupled to the center fre-
quency of the current measurement.
Remote Control - Commands