Remote Commands
R&S
®
Cable Rider ZPH
533
User Manual 1321.0950.02 ─ 07
larly since faulty commands from the controller to the instrument are recorded there as
well.
14.6.12.6
Reset Values of the Status Reporting System
Table 14-13
contains the different commands and events causing the status reporting
system to be reset. None of the commands, except
*RST
and
SYSTem:PRESet
, influ-
ences the functional instrument settings. In particular, DCL does not change the instru-
ment settings.
Table 14-13: Resetting the status reporting system
Event
Switching
on supply
voltage
DCL, SDC
Power-On-
Status-Clear
(Device
Clear,
Selected
Device
Clear)
*RST or
SYS-
Tem:PRE-
Set
STA-
Tus:PRESet
*CLS
Effect
0
1
Clear STB, ESR
-
yes
-
-
-
yes
Clear SRE, ESE
-
yes
-
-
-
-
Clear PPE
-
yes
-
-
-
-
Clear EVENt parts of
the registers
-
yes
-
-
-
yes
Clear ENABle parts
of all OPERation and
QUEStionable regis-
ters; Fill ENABle
parts of all other reg-
isters with "1".
-
yes
-
-
yes
-
Fill PTRansition parts
with "1"; Clear
NTRansition parts
-
yes
-
-
yes
-
Clear error queue
yes
yes
-
-
-
yes
Clear output buffer
yes
yes
yes
1)
1)
1)
Clear command pro-
cessing and input
buffer
yes
yes
yes
-
-
-
1) Every command being the first in a program message, i.e., immediately following a
<PROGRAM MESSAGE TERMINATOR> clears the output buffer.
14.6.12.7
Remote Commands of the Status Reporting System
The following commands control the status-reporting system
*RST
does not influence
the status registers.
Remote Control - Commands