Remote Commands
R&S
®
Cable Rider ZPH
384
User Manual 1321.0950.02 ─ 07
Suffix:
<1...2>
.
1...2
Parameters:
<Return values>
<0>
Not calibrated
<1>
Calibrated
Example:
CAL:MODE?
If current calibration state is not calibrated, a '0' is returned.
Usage:
Query only
CALibration<1...2>:STARt?
<CalibrationType>
This command selects a calibration method and initiates calibration.
After the command has been sent, it returns instructions on what to do to continue the
calibration process and finish it successfully.
The numeric suffix at CALibration is irrelevant for this command.
This command is a query and therefore has no *RST value.
Suffix:
<1...2>
.
1...2
Parameters:
<CalibrationType>
S11Cal | S11easy
S11Cal
Full 1-Port calibration
S11easy
Easy 1-Port calibration (Port 1) (CAT mode only)
<Return values>
<Connect Open to Port 1>
Connect the open to port 1 before resuming calibration.
<Connect Short to port 1>
Connect the short to port 1 before resuming calibration.
<Connect Load to port 1>
Connect the short to port 1 before resuming calibration.
<Disconnect DUT from port 1>
<Calibration done>
Calibration has been finished. No further actions required.
Example:
CAL:STAR? S11easy
Selects and initiates an Easy 1-Port calibration and would return
e.g.:
Calibration done
Calibration has been finished. No further actions required.
Usage:
Query only