Remote Commands
R&S
®
Cable Rider ZPH
396
User Manual 1321.0950.02 ─ 07
INPut:IMPedance
<Impedance>
This command selects the nominal input impedance. The set impedance is taken into
account in all level indications of results.
The setting 75 Ω should be selected, if the 50 Ω input impedance is transformed to a
higher impedance using a 75 Ω adapter of the RAZ type (= 25 Ω in series to the input
impedance of the instrument). The correction value in this case is 1.76 dB = 10 log (75
Ω / 50 Ω).
Parameters:
<Impedance>
50 | 75
*RST:
50 Ω
Default unit: Ohm
Example:
INP:IMP 75
Sets the input impedance to 75 Ohms.
[SENSe:]CORRection:TRANsducer<1...2>[:STATe]
<State>
This command turns a transducer factor on and off.
Before turning it on, you have to select a transducer factor with
[SENSe:
]CORRection:TRANsducer<1...2>:SELect
on page 396.
The numeric suffix at TRANsducer specifies the primary or secondary transducer.
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
CORR:TRAN1 ON
Activates the primary transducer.
[SENSe:]CORRection:TRANsducer<1...2>:SELect
<TransducerName>
This command selects a transducer factor.
If <name> does not exist yet, a new transducer factor is created.
The numeric suffix at TRANsducer specifies the primary or secondary transducer.
Parameters:
<TransducerName> String containing the filename of the transducer factor.
If the file does not exist, the R&S Cable Rider ZPH creates a
new transducer factor.
Example:
CORR:TRAN2:SEL 'FSH-Z38.sectrd'
Selects the FSH-Z38 secondary transducer factor.
[SENSe:]CORRection:TRANsducer<1...2>:UNIT?
This command queries the unit of the current transducer factor.
Remote Control - Commands