Remote Commands
R&S
®
Cable Rider ZPH
344
User Manual 1321.0950.02 ─ 07
[SENSe:]FREQuency:SETTings:COUPling:ENABle
<State>
This command couples or decouples the frequency settings of the DTF measurement
to the other Cable and Analyzer measurements.
Parameters:
<State>
ON | 1
Frequency settings are coupled. When you change the fre-
quency or span in the DTF measurement, they also change in
the other Cable and Analyzer measurements (reflection, trans-
mission and cable loss).
ON | 1
Frequency settings of the DTF measurement are independent.
When you change the frequency or span in the DTF measure-
ment, they do not change in the other VNA measurements
(reflection, transmission and cable loss).
*RST:
ON
Example:
FREQ:SETT:COUP:ENAB OFF
Decouple frequency and span.
CALCulate<1...2>:DTF:DISTance:STARt
<DtfDistanceStart>
This command defines the start distance of the cable measurement.
The numeric suffix at CALCulate is irrelevant for this command.
Suffix:
<1...2>
.
1...2
Parameters:
<DtfDistanceStart>
Numeric value in the range from 3 m (10 ft) to 1500 m (4921 ft)
The unit is either meter or feet, depending on your selection.
*RST:
3 m (10 ft)
Default unit: m
Example:
CALC:DTF:DIST:STAR 50m
Sets the starting point of the measurement to 50 m.
CALCulate<1...2>:DTF:DISTance:STOP
<DtfDistanceStop>
This command defines the end point of the cable measurement.
The numeric suffix at CALCulate is irrelevant for this command.
Suffix:
<1...2>
.
1...2
Remote Control - Commands