Spectrum Analyzer Mode
R&S
®
Cable Rider ZPH
238
User Manual 1321.0950.02 ─ 07
The selected marker is highlighted in the
Measurement Result View
and its corre-
sponding label in the trace window turns focus with a blue frame around the marker
label. The R&S Cable Rider ZPH opens a marker input field.
2. Select the "Delete Marker" softkey.
3. Select the "Delete Selected" menu item.
4. Confirm the selection by pressing the rotary knob.
The R&S Cable Rider ZPH deletes the marker.
Deactivating markers
If you delete marker 1 (M1), all delta markers that are relative to that marker are also
deleted.
Removing delta markers only
1. Select the delta marker you want to delete in the
Measurement Result View
.
Alternatively, press the [MARKER] key.
2. Select the "Delete Marker" softkey.
3. Select the "Delete All Delta" menu item.
4. Confirm the selection by pressing the rotary knob.
The R&S Cable Rider ZPH deletes all delta markers.
Removing all markers at the same time.
1. Press the [MARKER] key.
2. Select the "Delete Marker" softkey.
3. Select the "Delete All" menu item.
4. Confirm the selection by pressing the rotary knob.
The R&S Cable Rider ZPH deletes all markers and delta markers.
Special touchscreen gesture
Alternatively, draw an "x" on the trace window to delete all markers and delta markers
on the trace window.
See
Chapter 5.1.6, "Delete All Markers"
,
on page 92.
8.2.8.7
Using Marker Search Limits
The R&S Cable Rider ZPH allows you to use only a limited section of the trace for the
"Set to Peak", "Set to Next Peak" and "Minimum" functions.
1. Press the [MARKER] key.
Configuring Spectrum Measurements