Remote Commands
R&S
®
Cable Rider ZPH
349
User Manual 1321.0950.02 ─ 07
The numeric suffix at DISPlay is irrelevant for this command.
Suffix:
<1...2>
.
1...2
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
DISP:PHAS:UNWR ON
Activates the phase unwrap.
DISPlay<1...2>:VSWR:Y:SCALe:MINimum
<DisplayRange>
This command defines the bottom value of the vertical axis for the VSWR measure-
ment format.
The numeric suffix at DISPlay is irrelevant for this command.
Suffix:
<1...2>
.
1...2
Parameters:
<DisplayRange>
Numeric value in the range from 1.0 to 70.
*RST:
1.0
Example:
DISP:VSWR:Y:SCAL:MIN 3
Defines a bottom value of 3 for the vertical axis.
DISPlay<1...2>:VSWR:Y:SCALe:MAXimum
<DisplayRange>
This command defines the top value of the vertical axis for the VSWR measurement
format.
The numeric suffix at DISPlay is irrelevant for this command.
Suffix:
<1...2>
.
1...2
Parameters:
<DisplayRange>
Numeric value in the range from 1.1 to 71.
*RST:
21
Example:
DISP:VSWR:Y:SCAL:MAX 25
Defines a top value of 25 for the vertical axis.
DISPlay<1...2>:VSWR:Y:SCALe
<DisplayRange>
This command defines the bottom value of the vertical axis for the VSWR measure-
ment format.
The numeric suffix at DISPlay is irrelevant for this command.
Remote Control - Commands