![Rohde & Schwarz 1321.1211.02 User Manual Download Page 443](http://html1.mh-extra.com/html/rohde-and-schwarz/1321-1211-02/1321-1211-02_user-manual_1477772443.webp)
Remote Commands
R&S
®
Cable Rider ZPH
436
User Manual 1321.0950.02 ─ 07
OBANdwidth | OBWidth
Occupied Bandwidth measurement.
TDMA
TDMA power measurement.
Example:
CALC:MARK:FUNC:POW:SEL CPOW
Selects the channel power measurement.
CALCulate<1...2>:MARKer<1...6>:FUNCtion:POWer[:STATe]
<State>
This command turns a power measurement on and off.
You can select a power measurement with
CALCulate<1...2>:MARKer<1...6>:
FUNCtion:POWer:SELect
on page 435.
The numeric suffix at CALculate and MARKer are irrelevant for this command.
Parameters:
<State>
ON | OFF
ON
Turns the power measurement on.
OFF
Performs a basic spectrum measurement.
*RST:
OFF
Example:
CALC:MARK:FUNC:POW OFF
Switches off the power measurement.
CALCulate<1...2>:MARKer<1...6>:FUNCtion:POWer:PRESet
<Standard>
This command selects one of the predefined configurations for a telecommunications
standard. This command only works if you have turned on power measurements with
CALCulate<1...2>:MARKer<1...6>:FUNCtion:POWer[:STATe]
on page 436.
The configuration for a standard is in line with the specifications and includes parame-
ters like weighting filter, channel bandwidth and spacing, resolution and video band-
width, as well as detector and sweep time.
The numeric suffix at CALculate and MARKer are irrelevant for this command.
Parameters:
<Standard>
String containing the filename of the standard.
Example:
CALC:MARK:FUNC:POW:PRES '3GPP WCDMA.chpstd'
Selects the 3GPP WCDMA standard for channel power mea-
surements.
Usage:
Setting only