Remote Commands
R&S
®
Cable Rider ZPH
470
User Manual 1321.0950.02 ─ 07
[SENSe:]ADEMod:FM:OFFSet?
This command displays the Frequency Deviation value for FM.
Example:
ADEM:FM:OFFS?
Reads out the Frequency Deviation value for FM.
Usage:
Query only
CALCulate<1...2>:MARKer:FUNCtion:ADEMod:CARRier[:RESult]?
This command displays the Carrier Power value for both AM/FM.
The numeric suffix at CALCulate is irrelevant for this command.
Suffix:
<1...2>
.
1...2
Example:
CALC:MARK:FUNC:ADEM:CARR?
Reads out the Carrier Power value for both AM/FM.
Usage:
Query only
CALCulate<1...2>:MARKer:FUNCtion:ADEMod:AFRequency[:RESult]?
This command queries the modulation frequency.
The numeric suffix at CALCulate is irrelevant for this command
Suffix:
<1...2>
.
1...2
Example:
CALC:MARK:FUNC:ADEM:AFR?
Reads out the modulation frequency
Usage:
Query only
CALCulate<1...2>:MARKer:FUNCtion:ADEMod:AM[:RESult]?
<MeasType>
This command displays the results of frequency deviation for AM.
The numeric suffix at CALCulate is irrelevant for this command.
Suffix:
<1...2>
.
1...2
Parameters:
<MeasType>
PPEak | MPEak | MIDDle | RMS
PPEak
+Peak value
MPEak
-Peak value
MIDDle
-
±
Peak value
/
2
Remote Control - Commands