Basic Operation
R&S
®
Cable Rider ZPH
54
User Manual 1321.0950.02 ─ 07
Activating / Deactivating audio alert for battery low-level state
The R&S Cable Rider ZPH also allows you to turn on an audio signal that indicates
that the battery has reached its low-level state.
1. In the "Instrument Setup" dialog box, select the "Battery Low Level Beep" item.
2. Select either "Repetitive" or "Once" to turn on the audio signal.
If you have selected "Once", the R&S Cable Rider ZPH beeps once if the battery
runs out of power. For a continuous beep, select "Repetitive".
3. Select "Off" to turn off the beeper.
Auto power up
The R&S Cable Rider ZPH allows you to auto power up the instrument when supplying
with an AC power, without pressing the [POWER] key button.
1. In the "Instrument Setup" dialog box, select the "Auto Power Up on AC" menu item.
2. Select "On" to turn on the auto power up mode.
On the next power up cycle, the R&S Cable Rider ZPH goes into the auto power
up mode when supplying with an AC power.
4.2.9
Internal Alignment
R&S Spectrum Rider option
R&S ZPH-K35 (order number: 1321.1570.02) option is required to operate the R&S
Cable Rider ZPH internal alignment.
●
For model .12, both frequency clock and reference level adjustment are supported.
Reference level adjustment is only possible if R&S ZPH-K1 (order number:
1334.5604.02) is available in the R&S Cable Rider ZPH.
●
For model .02, only frequency clock adjustment is supported.
Configuring the R&S Cable Rider ZPH