Remote Commands
R&S
®
Cable Rider ZPH
369
User Manual 1321.0950.02 ─ 07
Example:
See
CALCulate<1...2>:MARKer<1...6>:FUNCtion:
NOISe:RESult?
on page 369.
CALCulate<1...2>:MARKer<1...6>:FUNCtion:NOISe:RESult?
This command queries the result of the noise measurement.
To get a valid result, you have to perform a complete sweep with synchronization to the
sweep end before reading out the result. This is only possible in single sweep mode.
This command is an event and therefore has no *RST value and no query.
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
MARKer selects the marker.
Example:
INIT:CONT OFF
Turns on single sweep mode.
CALC:MARK2 ON
CALC:MARK2:FUNC:NOIS ON
Turns on marker 2 and assigns the noise measurement to that
marker.
INIT;*WAI
CALC:MARK2:NOIS:RES?
Performs the measurement and queries the noise marker
results.
Usage:
Query only
CALCulate<1...2>:MARKer<1...6>:FUNCtion:REFerence
This command matches the reference level to the power level of a marker.
If you use a delta marker, the R&S Cable Rider ZPH turns it into a normal marker.
This command is an event and therefore has no *RST value and no query.
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
MARKer selects the marker.
Example:
CALC:MARK1:FUNC:REF
Matches the reference level to the power level of marker 1.
Usage:
Event
14.6.2.7
Using Display Lines and Limit Lines
The following commands define the display lines and limit lines.
Display Lines
The following commands define the position of the display line.
Remote Control - Commands