Remote Commands
R&S
®
Cable Rider ZPH
408
User Manual 1321.0950.02 ─ 07
CALCulate<1...2>:MATH<1...2>:STATe
<State>
This command turns trace mathematics on and off.
The numeric suffix at CALCulate is irrelevant for this command.
The numeric suffix at MATH selects the number of the trace.
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
CALC:MATH:STAT ON
Switches on the trace mathematics.
DISPlay<1...2>[:WINDow]:TRACe<1...2>[:STATe]
<State>
This command turns a trace on and off.
The numeric suffix at DISPlay is irrelevant for this command.
The numeric suffix at TRACe selects the number of the trace.
Parameters:
<State>
ON | OFF
*RST:
ON for TRACe1, OFF for TRACe2
Example:
DISP:TRAC2 ON
Turns on the trace 2.
DISPlay<1...2>[:WINDow]:TRACe<1...2>:MEMory<1...3>[:STATe]
<State>
This command turns the memory trace on and off.
The numeric suffix at DISPlay and MEMory are irrelevant for this command.
The numeric suffix at TRACe selects the number of the trace.
Suffix:
<1...2>
.
1...2
<1...2>
1...2
<1...3>
1...3
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
DISP:TRAC:MEM ON
Activates the memory trace.
DISPlay<1...2>[:WINDow]:TRACe<1...2>:MODE
<DisplayMode>
This command selects the trace mode.
Remote Control - Commands