Remote Commands
R&S
®
Cable Rider ZPH
393
User Manual 1321.0950.02 ─ 07
DISPlay<1...2>[:WINDow]:TRACe<1...2>:Y[:SCALe]:ADJust
This command automatically scales the vertical axis for optimum display results.
This command is an event and therefore has no query and no *RST value.
The numeric suffix at DISPlay and TRACe are irrelevant for this command.
Example:
DISP:TRAC:Y:ADJ
Adjusts the y-axis.
Usage:
Event
DISPlay<1...2>[:WINDow]:TRACe<1...2>:Y:SPACing
<ScalingType>
This command selects the scaling type of the vertical axis.
The numeric suffix at DISPlay and TRACe are irrelevant for this command.
Parameters:
<ScalingType>
LINear | LOGarithmic
LINear
Selects a linear scale (%).
LOGarithmic
Selects a logarithmic scale.
*RST:
LOGarithmic
Example:
DISP:TRAC:Y:SPAC LIN
Selects linear scaling of the level axis.
DISPlay<1...2>[:WINDow]:TRACe<1...2>:Y[:SCALe]
<DisplayRange>
This command defines the display range of the vertical axis.
Note that you have to set a logarithmic scaling before you can use this command with
DISPlay<1...2>[:WINDow]:TRACe<1...2>:Y:SPACing
on page 393. For a lin-
ear scale, you cannot modify the display range as it is fixed.
The numeric suffix at DISPlay and TRACe are irrelevant for this command.
Parameters:
<DisplayRange>
Range:
1 dB to 150 dB
*RST:
100 dB
Default unit: dB
Example:
DISP:TRAC:Y 110dB
Sets the display range to 110 dB.
DISPlay<1...2>[:WINDow]:TRACe<1...2>:Y[:SCALe]:RLEVel
<RefLevel>
This command defines the reference level.
Remote Control - Commands