Remote Commands
R&S
®
Cable Rider ZPH
494
User Manual 1321.0950.02 ─ 07
List of commands
CALCulate<1...2>:PMETer:RELative[:MAGNitude]
.............................................................494
CALCulate<1...2>:PMETer:RELative[:MAGNitude]:AUTO
...................................................494
CALCulate<1...2>:PMETer:RELative[:MAGNitude]:OFFSet
................................................ 494
UNIT<1...2>:PMETer:POWer
...........................................................................................494
CALCulate<1...2>:PMETer:RELative[:MAGNitude]
<RefLevel>
This command sets the reference value for relative measurements.
The numeric suffix at CALCulate is irrelevant for this command.
Parameters:
<RefLevel>
Default unit: dBm
Example:
CALC:PMET:REL 30
The reference value to 30 dBm.
CALCulate<1...2>:PMETer:RELative[:MAGNitude]:AUTO
<ONCE>
This command sets the current measurement result as the reference level for relative
measurements.
This command is an event and therefore has no *RST value and no query.
The numeric suffix at CALCulate is irrelevant for this command.
Parameters:
<ONCE>
ONCE
Example:
CALC:PMET:REL ONCE
CALCulate<1...2>:PMETer:RELative[:MAGNitude]:OFFSet
<RefLvlOffset>
This command sets an offset for the reference value.
The numeric suffix at CALCulate is irrelevant for this command.
Parameters:
<RefLvlOffset>
Default unit: dB
Example:
CALC:PMET:REL -10
UNIT<1...2>:PMETer:POWer
<Unit>
This command selects the unit of the power sensor.
The numeric suffix at UNIT has the following effects:
Table 14-4: Power measurement with R&S FSH-Z1, R&S FSH-Z18 and USB power sensors:
Unit 1
Power unit.
Unit 2
Not available.
Remote Control - Commands