Spectrum Analyzer Mode
R&S
®
Cable Rider ZPH
252
User Manual 1321.0950.02 ─ 07
If you are using a transducer with the unit dB
µ
A/m, it is not possible to select
another unit. Level indication is entirely in dB
µ
A/m.
8.4.2
Setting the Reference Level
The transducer shifts the trace by its value as a function of frequency. Positive trans-
ducer values increase the level, negative values reduce it. To ensure that the trace is
always within the diagram, the R&S Cable Rider ZPH adjusts the reference level
accordingly. The reference level is shifted by the maximum transducer value in the
positive or negative direction.
8.4.3
Frequency Range of Transducer
If the set frequency range is wider than the span in which a transducer is defined, the
R&S Cable Rider ZPH assumes the transducer values outside the defined range to be
zero.
8.4.4
Data Sets Containing Transducer Factors
The R&S Cable Rider ZPH stores data sets together with any transducer factors that
may have been active for the measurement in question. When such a dataset is recal-
led, the associated transducer factor(s) are switched on as well. Transducer factors
recalled as part of a data set do however not appear in the list of transducer factors.
Using Transducer Factors