![Rohde & Schwarz 1321.1211.02 User Manual Download Page 165](http://html1.mh-extra.com/html/rohde-and-schwarz/1321-1211-02/1321-1211-02_user-manual_1477772165.webp)
Cable and Antenna Test Mode
R&S
®
Cable Rider ZPH
158
User Manual 1321.0950.02 ─ 07
Figure 7-10: Screen layout with active markers
1 =
Measurement Result View
2 = Marker (solid line)
3 = Delta marker (dotted line)
4 = Active marker label (see highlighted line on the marker list as well as the marker label )
5 = Marker label: M(x)
6 = Delta marker label: D(x)
7 = Marker input field
8 = Marker menu
9 =
Parameter View
Positioning Markers
Special touchscreen gesture
Double tap on the touchscreen to position new marker on the trace window. The first
marker that is positioned on the trace window is always the main marker, the following
markers added on the trace window are the delta markers.
Tab and drag on the marker icon to change the marker position.
See
Chapter 5.1.4, "Add Marker"
,
on page 91 and
Chapter 5.1.5, "Move Marker"
,
on page 91.
1. Press the [MARKER] key.
If no marker has been activated, the R&S Cable Rider ZPH automatically activates
the main marker and positions it on the maximum level that has been measured. In
addition, the marker frequency input field opens.
You can perform the following actions:
● Tab and drag on the marker icon to change the marker position.
● Position the marker with the rotary knob
Analyzing Measurement Results