Using the Internal Power Meter
R&S
®
Cable Rider ZPH
265
User Manual 1321.0950.02 ─ 07
Defining the channel bandwidth for model .12
For model .12, the R&S Cable Rider ZPH provides additional setting to select the
channel bandwidth.
1. Press the [MEAS] key.
2. Select the "Channel BW" softkey.
The R&S Cable Rider ZPH opens an input field to define the channel bandwidth.
3. Enter the channel bandwidth you require.
The R&S Cable Rider ZPH performs a measurement on the selected channel.
Note that it is not possible to change the measurement time, resolution bandwidth
and frequency span.
Activating signal source for model .02
For model .02, the R&S Cable Rider ZPH provides the setting to enable or disable the
built-in signal source in the power meter mode.
Once the signal source is enabled, you can output a signal source of -10 dBm at the
designated frequency from the RF output connector.
► Select the "SigGen On" softkey.
The R&S Cable Rider ZPH output a signal source of -10 dBm at the RF output con-
nector.