Remote Commands
R&S
®
Cable Rider ZPH
482
User Manual 1321.0950.02 ─ 07
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
CALC:DELT3 ON
Turns on delta marker 3 or turn marker 3 into a delta marker.
CALCulate<1...2>:DELTamarker<1...6>:AOFF
This command turns off all active delta markers.
This command is an event and therefore has no query and no *RST value.
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
DELTamarker selects the deltamarker.
Example:
CALC:DELT:AOFF
Turns off all delta markers.
Usage:
Event
CALCulate<1...2>:DELTamarker<1...6>:X
<FrequencyOrTime>
This command positions a delta marker on a particular coordinate on the horizontal
axis.
Note that it is possible to place the marker outside the visible trace. In that case, this
value is invalid.
If necessary, the corresponding delta marker is activated first.
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
DELTamarker selects the deltamarker.
Parameters:
<FrequencyOrTime> Numeric value that indicates the coordinate on the horizontal
axis.
Range:
Maximum span.
Default unit: GHz
Example:
CALC:DELT:MOD REL
Delta marker positions are relative to marker 1.
CALC:DELT2:X 10.7MHz
Positions delta marker 2 10.7 MHz to the right of marker 1.
CALC:DELT2:X?
CALC:DELT2:X:REL?
Queries the absolute and relative position of delta marker 2.
CALCulate<1...2>:DELTamarker<1...6>:X:RELative
<FrequencyOrTime>
This command positions a delta marker on a position relative to the reference marker.
If necessary, the corresponding delta marker is activated first.
Remote Control - Commands