Remote Commands
R&S
®
Cable Rider ZPH
348
User Manual 1321.0950.02 ─ 07
Parameters:
<DisplayRange>
Numeric value in the range from 1 dB to 150 dB.
The number you enter is rounded up to the next possible display
range. For example, if you enter 9, the R&S Cable Rider ZPH
automatically sets the display range to 10.
*RST:
100 dB
Default unit: dB
Example:
DISP:MAGN:Y:SCAL 50 DB
Defines a display range of 50 dB.
DISPlay<1...2>:MAGNitude:Y:SPACing
<ScalingType>
This command selects the scaling of the vertical axis for the magnitude measurement
format.
The numeric suffix at DISPlay is irrelevant for this command.
Suffix:
<1...2>
.
1...2
Parameters:
<ScalingType>
LINear | LOGarithmic
LINear
Selects a linear scale (%).
LOGarithmic
Selects a logarithmic scale (dB).
*RST:
LOGarithmic
Example:
DISP:MAGN:Y:SPAC LIN
Selects linear scaling.
DISPlay<1...2>:PHASe:REFerence
<RefLevel>
This command sets the reference value for the phase measurement format.
The numeric suffix at DISPlay is irrelevant for this command.
Suffix:
<1...2>
.
1...2
Parameters:
<RefLevel>
Numeric value in the range from -100000° to 100000°.
*RST:
0 dB
Example:
DISP:MAGN:REF -10
Sets the reference level to -10 dB
DISPlay<1...2>:PHASe:UNWRap
<State>
This command removes the restriction limiting the value range to +/- 180
°
.
Remote Control - Commands