Remote Commands
R&S
®
Cable Rider ZPH
475
User Manual 1321.0950.02 ─ 07
DISPlay<1...2>[:WINDow]:TRACe<1...2>:Y[:SCALe]:PDIVision
<DevPerDiv>
This command sets the Y scale (deviation per division) in Hz (FM) or percent (AM).
The numeric suffix at DISPlay and TRACe is irrelevant for this command.
Suffix:
<1...2>
.
1...2
<1...2>
1...2
Parameters:
<DevPerDiv>
Numeric value with the unit range of Hz | kHz| MHz| MAHz| GHz
| PCT.
Default unit: GHz
Example:
DISP:TRAC:Y:PDIV 5 PCT
Set the Y scale (deviation per division) as 5 % in the AM modu-
lation.
DISPlay<1...2>[:WINDow]:TRACe<1...2>:Y[:SCALe]:RPOSition
<RefLvlPosition>
This command defines the position of the reference level on the display grid.
First, you have to set a logarithmic scale for the vertical axis with
DISPlay<1...2>[:
WINDow]:TRACe<1...2>:Y:SPACing
on page 393. For a linear scale, you cannot
modify the reference position.
The numeric suffix at DISPlay and TRACe are irrelevant for this command.
Parameters:
<RefLvlPosition>
Defines the reference position in %.
Range:
0 to 10
*RST:
10
Example:
DISP:TRAC:Y:RPOS 5
Sets the reference position to the 5th grid line.
DISPlay<1...2>[:WINDow]:TRACe<1...2>:Y[:SCALe]:RLEVel
<RefLevel>
This command defines the reference level.
With a reference level offset ≠ 0, the value range of the reference level is modified by
the offset. You can set the offset with
DISPlay<1...2>[:WINDow]:
TRACe<1...2>:Y[:SCALe]:RLEVel:OFFSet
on page 394.
The numeric suffix at DISPlay and TRACe are irrelevant for this command.
Parameters:
<RefLevel>
Sets the reference level; the unit depends on
UNIT<1...2>:
POWer
on page 397.
Range:
Specified in the data sheet.
*RST:
–20 dBm
Default unit: dBm
Remote Control - Commands