Remote Commands
R&S
®
Cable Rider ZPH
365
User Manual 1321.0950.02 ─ 07
Example:
INIT:CONT OFF
CALC:MARK ON
Turns on single sweep mode and marker 1.
CALC:MARK:COUN ON
Turns on the frequency counter for marker 1.
INIT;*WAI
CALC:MARK:COUN:FREQ?
Performs a measurement and queries the results of the fre-
quency counter.
CALCulate<1...2>:MARKer<1...6>:FREQuency:MODE
<Mode>
This command selects the marker frequency display mode.
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
MARKer selects the marker.
Parameters:
<Mode>
FREQuency | CHANnel
FREQuency
Sets the marker frequency mode to frequency input (in Hz).
CHANnel
Sets the marker frequency mode to channel input (as a channel
number).
*RST:
FREQ
Example:
CALC:MARK:FREQ:MODE FREQ
Selects the frequency display mode.
CALCulate<1...2>:MARKer<1...6>:FUNCtion:CENTer
This command matches the center frequency to the frequency of a marker.
If you use a delta marker, the R&S Cable Rider ZPH turns it into a normal marker.
This command is an event and therefore has no *RST value and no query.
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
MARKer selects the marker.
Example:
CALC:MARK2:FUNC:CENT
Matches the center frequency to the frequency of marker 2.
Usage:
Event
CALCulate<1...2>:MARKer<1...6>:FUNCtion:DEModulation[:STATe]
<State>
This command turns the audio demodulator on and off when the measurement hits a
marker position.
Remote Control - Commands