![Rohde & Schwarz 1321.1211.02 User Manual Download Page 390](http://html1.mh-extra.com/html/rohde-and-schwarz/1321-1211-02/1321-1211-02_user-manual_1477772390.webp)
Remote Commands
R&S
®
Cable Rider ZPH
383
User Manual 1321.0950.02 ─ 07
Usage:
Event
CALibration<1...2>:ATTenuation:STATus?
This command queries the current calibration state.
The numeric suffix at CALibration is irrelevant for this command.
This command is a query and therefore has no *RST value.
Suffix:
<1...2>
.
1...2
Parameters:
<Return values>
<NORMalized>
Full calibration.
<APPRoximate>
Approximate calibration (caused by a change of attenuation, for
example): measurement uncertainty must be anticipated
Example:
CAL:ATT:STAT?
Queries the calibration status of the R&S Cable Rider ZPH.
Usage:
Query only
CALibration<1...2>:CONTinue?
This command resumes calibration after a calibration standard has been connected.
After the command has been sent, it returns instructions on what to do to continue the
calibration process and finish it successfully.
The numeric suffix at CALibration is irrelevant for this command.
This command is a query and therefore has no *RST value.
Suffix:
<1...2>
.
1...2
Example:
CAL:CONT?
Resumes calibration and would return, e.g.:
'Connect Load to Port 1'
Connect the load before resuming calibration.
Usage:
Query only
CALibration<1...2>:MODE?
This command queries the general current calibration state.
The numeric suffix at CALibration is irrelevant for this command.
This command is a query and therefore has no *RST value.