Spectrum Analyzer Mode
R&S
®
Cable Rider ZPH
169
User Manual 1321.0950.02 ─ 07
Figure 8-1: Screen layout for channel power measurements
1 = Measurement mode
2 = Channel power numerical
3 = Measurement standard
4 = Channel bandwidth
5 = Channel bandwidth graphical (blue lines)
6 =
Parameter view
7 =
Config Overview
8 = Channel power measurement softkey menu
8.1.2.1
Selecting the Standard
If you need to perform measurements that are conformed to a telecommunications
standard, you can activate one of the predefined standards that are already stored in
the R&S Cable Rider ZPH memory. However, you can create new configurations to
perform measurements on other standards as well.
1. Press the [MEAS] key.
2. Select the "Standard" softkey.
The R&S Cable Rider ZPH opens a file manager dialog box to select the standard.
3. Select one of the available standards.
The R&S Cable Rider ZPH loads the configuration of the selected standard. It
automatically sets the optimal span, resolution bandwidth, video bandwidth, sweep
time and detector for the standard.
You can create and edit standards with the R&S InstrumentView software and transfer
them to the R&S Cable Rider ZPH via the USB or the LAN interface. The number of
standards the R&S Cable Rider ZPH can store in its memory depends on the number
of other data sets stored on the R&S Cable Rider ZPH.
Performing Spectrum Measurements