
Remote Commands
R&S
®
Cable Rider ZPH
432
User Manual 1321.0950.02 ─ 07
This command is an event and therefore has no *RST value and no query.
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
LIMit selects the limit line.
Suffix:
<1...2>
.
1...2
<1...2>
1...2
Parameters:
<LimitLine>
String containing the filename of the upper limit line.
Example:
CALC:LIM1:UPP:SEL 'GSM_Upper.rellim'
Selects the upper limit line.
Usage:
Setting only
CALCulate<1...2>:LIMit<1...2>:LOWer:THReshold
<Threshold>
This command defines the level of a lower threshold limit line.
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
LIMit selects the limit line.
Suffix:
<1...2>
.
1...2
<1...2>
1...2
Parameters:
<Threshold>
Numeric value whose unit depends on the unit you have cur-
rently selected for the vertical axis.
Default unit: dBm
Example:
CALC:LIM1:LOW:THR -10DBM
Defines a threshold of -10 dBm.
CALCulate<1...2>:LIMit<1...2>:UPPer:THReshold
<Threshold>
This command defines the level of an upper threshold limit line.
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
LIMit selects the limit line.
Suffix:
<1...2>
.
1...2
<1...2>
1...2
Parameters:
<Threshold>
Numeric value whose unit depends on the unit you have cur-
rently selected for the vertical axis.
Default unit: dBm
Remote Control - Commands