3 - 36
(iii) Upon STOP during gradient elution
①
Initial
Liquid is delivered in the initial status (initial condition of
program No. 1 in this example).
②
Gradient
Gradient elution is carried out (program No. 1 is executed
in this example).
③
Initial
Liquid is delivered in the initial status (initial condition of
program No. 1 in this example).
④
Gradient
Gradient elution is carried out (program No. 1 is executed
in this example). In this case, the step stopped halfway
will be carried out.
TIME %A %B %C %D FLOW PRESS
[
IN
]
0.0
10 90
0
0 1.000
5.5 C 1
TIME %A %B %C %D
FLOW PRESS
[
GR
]
10.0 20 80 0 0 1.000 5.5 C
1
TIME %A %B %C %D
FLOW PRESS
[
LC
]
0.0 10 90 0 0 1.000 5.5 C
2
TIME %A %B %C %D
FLOW PRESS
[
IN
]
10.0 20 80 0 0 1.000 5.5 C
1