<Utility setting screen>
UTILITY (OFFSET=1, TIME CONSTANT=2,
_5
USE PROG=3, PMT VOLT=4, OTHERS=5)
Press numeric key 5 and then the ENT key.
(b) The others setting screen appears.
<Others setting screen>
OTHERS (Em BANDWIDTH=1, ERROR OUT=2,
4
LEAK SENSOR=3, LAMP=4)
Press numeric key 4 and then the ENT key.
(c) The Xe lamp setting screen appears.
<Xe lamp setting screen>
Xe LAMP ON/OFF (OFF=0, ON=1)
0
Press numeric key 0 and then the ENT key.
(d) The
1
st
monitor screen appears, and OFF is indicated
at LAMP.
<1
st
monitor screen>
TIME Ex Em FL PROG LAMP [I]
0.0 250
350 0.0
1
OFF
(e) The xenon lamp is thus turned off. Although the lamp
is off, the fan continues rotating so as to cool down the
lamp. Wait for at least 30 minutes in this status
without turning off the power supply.
5 - 4