Shown below is the assembling method of tubing for 1-line pre-
heating, for example.
(1) Loosen the two hexagon bolts fastening the pre-heating
plate where the pre-heating tubing is to be installed, using
the hexagon screw wrench attached.
Appended Fig. 1-9 Disassembling of Pre-heating Plate
(2) Prepare a tube of approx. 800 mm long. Excluding the
portions on both ends (100 mm on the column side and
500 mm on the autosampler connected side), install the
remainder of the tube into the grooves in the pre-heating
base.
Column connected
side
100 mm
Pre-heating base
Groove
Autosampler
connected side
500 mm
Appended Fig. 1-10 Assembling of Tube
APPENDIX 1 - 14