(4) Detach the injection valve seal from the unit.
Unit
Injection valve seal
Injection valve stator
Stator setscrew
Stator ring
Fig. 5-18 Removal of Syringe Valve Stator
(5) In the reverse order of removal, attach a new syringe valve
seal, the stator ring, and the syringe valve stator. Then
tighten the stator setscrews. Secure these setscrews in
turn and by equal increments.
(6) In the reverse order of removal, connect the tubes and
pipes to the syringe valve.
(7) After replacement of the syringe valve seal, check that
there are no liquid leaks.
NOTES
: 1. While the inside of the valve is dry, avoid moving
the valve unless otherwise necessary. Otherwise,
the plunger seal will sear out abnormally.
2. If the valve is not filled with liquid, it lacks
lubricity. Therefore, the valve may not move
normally.