3.2 UTLTY Setting
3.2.1 Zero Pressure Calibration
The zero pressure point may change a little due to a secular
change of the pressure sensor. Please check the zero
pressure point once a month (Stop the pump, and then open the
drain valve). Refer to the following operation for calibration.
Monitor screen
PRESS is displayed as 0, and the zero pressure calibration is
completed.
UTLITY (CONDITION=1, UNIT=2, FLUSH=3
1 EVENT=4, LINK=5, P.WASH=6, OTHERS=7)
CONDITION SET (Z-ADJ=1, MOTOR-CTRL=2
1 PURGE=3, CALIBRATION=4, LPGR=5)
Select CONDITION
1 ENT
PRESSURE ZERO ADJUT (YES=1, NO=0)
1 (NOTE: OPEN DRAIN VLAVE)
Select Z-ADJ (ZERO ADJUST)
1 ENT
Select YES
1 ENT
Monitor screen
SET UP
UTLTY