3 - 41
3.3.2 Plunger Wash Pump Logbook
The total running time after replacement of a plunger wash
pump (option) is indicated. This function is enabled only when
the plunger wash pump is connected. This function is used as
an index of a replacement for plunger wash pump.
GLP (LOGBOOK=1, PRESSURE M=2, KEY LOCK=3)
1
LOGBOOK (PUMP SEAL=1, P.WASH=2,
2 A.VALVE=3, D.MIX=4, P.VALVE=5)
Select LOGBOOK
1 ENT
LOGBOOK (REPORT=1, RESET=2)
1
Select P.WASH
2 ENT
Monitor screen
GLP
Monitor screen
Select REPORT
1 ENT
Select RESET
2 ENT
P.WASH PUMP CHANGE (MM DD YYYY)
<2000 - 2100> 9 1 2010
Reset logbook
NOTE
: Enter the plunger wash pump replacing date.
Enter Month, Date, Year. And press ENT.
In other words, it can’t be reset without pressing the ENT key.
Even if the previous data is still indicated, reenter a desired
date and press ENT . The total running time will be indicated
as zero, and integration will be started.
Monitor screen
P.WASH TIME LAST CHANGED DATE
99999 h 09/01/2010
The total running time after plunger wash pump
replacement displays “h”.
ENT or ESC