5 - 1
5. PARTS REPLACEMENT
The following parts are consumables. Always prepare them for
replacement. Quantity of use is on a basis of rough standard,
assuming that the working days are 240 days a year, the working
hours are 8 hours a day. Normal life means usable time of the
part when the instrument is used with methanol flowing
(1.0 mL/min). However, since the values vary greatly depending
on the conditions (solvent, flow rate, pressure), prepare enough
quantity of each part, considering the actual conditions of use.
5.1 Pump Main Unit
Part No.
Name
Shape and Specifications
Position Used
Normal
Life
891-1213 Plunger
seal
(Pump seal)
Color: Black
Pump head for
Chromaster 5160
(4 months)
640 h
885-1884 Wash
mechanism seal
Wash mechanism
plunger guide
(4 month)
640 h
810-1163 Plunger
Pump head for
Chromaster 5160
1 - 3 year
885-2920 Packing
(Purge valve
seal)
Purge valve
(1 year)
1920 h
810-1128 Packing
Purge valve
(1 year)
1920 h
891-1228 Packing(PEEK)
Pressure sensor
holder
(1 year)
1920 h
Packing