5.2.1
5.2 Parts Replacement Method
5.2.1 Fuse
Replacement
WARNING
Electric Shock due to Contact with Inside of
Instrument
When replacing the fuse with a new one, be careful not to
receive an electric shock. To prevent this, be sure to turn
off the power switch before proceeding to fuse replacement.
(1) Item to Prepare
New fuse (P/N J821349)
Make sure to use a time lag fuse of the specified rating
(T5AL for 250 V).
(2) Check
Items
Make sure the power switch is OFF and the power cord is
disconnected from the instrument prior to fuse replacement.
(3) Fuse
Replacement
(a) Detach the fuse holder by pressing the knobs at both
sides.
Knobs
C941322
Fuse holder
Fig. 5-1 Fuse Replacement
(b) Replace the fuse and then re-attach the fuse holder by
pressing it.