(c) The following display will be presented.
<Others setting screen>
OTHERS (Em BANDWIDTH=1, ERROR OUT=2,
2
LEAK SENSOR=3, LAMP=4)
Select ERROR OUT.
Press numeric key 2 and then the ENT key.
(d) The following display will be presented.
<Error out signal selecting screen>
ERROR OUT SIGNAL (DISABLE=0, ENABLE=1)
1
Invalid (0) or valid (1) status of error out signal output is
settable.
Press numeric key (0 or 1) and then the ENT key.
* Default value: 1
(e) Error out setting is now completed and the 1
st
monitor
screen returns.
(3) LEAK SENSOR setting
Whether to use the liquid leak sensor or not is settable.
(a) Press
the
UTLTY
SET UP
key.
(b) The following display will be presented.
<Utility initial screen>
UTILITY (OFFSET=1, TIME CONSTANT=2,
_5
USE PROG=3, PMT VOLT=4, OTHERS=5)
Select OTHERS.
Press numeric key 5 and then the ENT key.
3 - 14