3.2.5
5) Upon completion of LAMP setting, the first monitor
screen reappears.
(e) Setting
the
Gain
Gain is selectable so as to match the flow cell to be used.
1) Press key.
SET UP
UTLTY
2) The utility setup screen appears.
Utility Setup Screen
UTILITY (OFFSET=1, TIME CONSTANT=2,
1
USE PROG=3, POLARITY=4, OTHERS=5)
Select OTHERS.
Press [numerical] key [5], and then the [ENT] key.
3) The OTHERS setup screen appears.
OTHERS (PROCESSOR=1, ERROR OUT=2
1
LEAK SENSOR=3, LAMP=4, GAIN=5)
Select GAIN.
Press [numerical] key [3], and then the [ENT] key.
4) The GAIN setup screen appears.
GAIN (NORMAL=0, MICRO=1)
0
Set the NORMAL or MICRO status of GAIN.
Press [numerical] key ([0] or [1]), and then the [ENT]
key.
* Default value: 1
5) Upon completion of GAIN setting, the first monitor
screen reappears.
3 - 19