5. SPARE PARTS
5.1 Replacement Parts and Consumables
Table 5-1 lists the replacement parts and consumables for this
instrument. It is suggested that a suitable quantity of each item
be stocked in the laboratory to minimize down time. For many
of these items (e.g. lamps), the quantity is dependent on the
level of use of the instrument.
The estimated service life of each part is based on the
assumption that the instrument is operated 160 hours/month
(8 hours/day), 1920 hours/year (240 days).
Table 5-1 Replacement Parts and Consumables
Part Number
Part Name
Estimated Service Life
Remarks
650-1500
(J851152)
150 W Xenon
lamp
Guaranteed life : 150 hours
Limit :
500
hours
Warranty period : 6 months after
delivery to
customer
890-2019 Long-life
xenon
lamp (Special
accessory)
Guaranteed life : 500 hours
Limit :
1000
hours
Warranty period : 6 months after
delivery to
customer
890-2662
Mercury lamp
Guaranteed life: 100 hours/
2 years
J821349
Time-lag fuse
5 A
890-2932 Flow
cell
(12
μ
L) 9600
hours/5
years
890-2684 Drain
tube
ID 0.33
×
0.62 T
9600 hours/5 years
Length: 2 m
Outside diameter:
1.57 mm
893-0833 Inlet
tube
ID 0.25
×
0.66 T
9600 hours/5 years
Length: 60 mm
Outside diameter:
1.57 mm (inlet
ferrule included)
L369053
Ferrule front
9600 hours/5 years
Part for connecting column
and inlet tube
Material: Daiflon
893-0825 Fitting
9600
hours/5
years
Example use
:
Column outlet
connection
For outer diameter :
1.57 mm
893-0826 Connecting
pipe
(SUS)
9600 hours/5 years
Example use:
Piping between
column outlet and
inlet tube
For thermostatic flow cell
Length :
350
mm,
Inner diameter : 0.25 mm,
Outer diameter : 1.57 mm,
Material :
SUS
5 - 1