5.2.1
(4) Operation
Check
Plug the power cord into the instrument and turn on the
power switch. Make sure the lamp comes on.
NOTE
: The
D
2
lamp does not come on if it has been
mounted improperly.
(5) On the monitor screen 2, check the wavelength (WL) and
reference energy (REF), and write down their values as a
record of the D
2
lamp after use. This record can be used
as a reference for the next D
2
lamp replacement and as a
historical record of D
2
lamp replacement.
Monitor Screen 2
250 191000 202000 ON
WL
SAMP
REF
CELL
[I]
(6) Enter the lamp replacement date as follows.
(a) Press key, and the following screen will appear.
GLP
1
D2 CHECK=3, W CHECK=4, Hg CHECK=5)
GLP (LOGBOOK=1, KEY LOCK=2,
(b) Press [numerical] key [1], and then the [ENT] key.
The following screen appears.
1
LAMP SELECT (D2=1, W=2, Hg=3)
(c) Press [numerical] key [1], and then the [ENT] key.
The following screen appears.
1
LOGBOOK (REPORT=1, RESET=2)
(d) Press [numerical] key [2], and then the [ENT] key.
The following screen appears.
<1–12> _9
15
2010
D2 LAMP CHANGE (MM DD YYYY)
5 - 7