3 - 4
Table 3-3 2
nd
Monitor Screen
Indication
Description
P-UNIT
Pressure unit (bar, psi, kgf/cm2, MPa can be
selected). Use UTLTY to select a pressure
unit.
P.MAX
Maximum operating pressure.
When the pressure becomes higher than the
setup value, a pump is stopped
automatically.
P.MIN
Minimum operating pressure
When the pressure becomes lower than the
setup value, a pump is stopped
automatically.
D.MIX Dynamic
mixer
USE/No USE
P.WASH Wash
pump
ON: Running, OFF: Suspended
NOTE
:
For Chromaster 5160 Pump, an internal maximum
operating pressure will be set up according to the flow
rate setting.
Pressure limit range:
0 to 2.500 mL/min ............ 60.0 MPa,
2.501 to 5.000 mL/min ..... 30.0 MPa.
(d) 3
rd
Monitor Screen
Table 3-4 3
rd
Monitor Screen
Indication Description
LOW GR.
Low pressure gradient unit connection
state
P.WASH
Plunger wash pump connection state
A.VALVE
Auto purge valve connection state
D.MIX
Dynamic mixer connection state
OPTION LOW GR. P.WASH A.VALVE D.MIX [ -- ]
YES YES YES YES 1