(continued)
4 ENT
Sample injection timing can be synchronized with pump
action. Reproducibility on retention time can be
enhanced.
NOTES
:
1. For using this function, it is required to
combine the autosampler with the
Chromaster 5110 pump. If the autosampler
is combined with any other model of pump,
the INJECTION TIMING error will be issued.
Take care not to incur this condition.
Since synchronization is made with the cam
position of the pump, the time period to be
taken before injection may be longer than
that in ordinary analysis.
2. Indicated below are the proper minimum flow
rate ranges for the respective pumps in
“USE” setting of this function. In cases
where a lower flow rate is selected, do not
specify “SYNCHRONIZED WITH PUMP
(PASS)”.
Model
Proper Minimum Flow
Rate Range (mL/min)
Chromaster 5110
pump
0.200
1 ENT
INJ TIMING (NO USE=0, USE=1)
0
Monitor screen reappears.
3 - 23