4 - 6
(7) Data
Check
Determine the coefficient of variation (RSD) of peak
retention time by the following calculations.
Mean value
n
xi
x
n
1
i
Standard deviation
1
n
)
x
xi
(
SD
n
1
i
2
Relative standard deviation
RSD
100(%)
x
SD
xi: Retention time due to each injection
n: Number of injections
Make sure the RSD value is 0.075% or less.
However, this value represents the precision of not the
pump alone but the entire chromatographic system.
Therefore, the value varies with each stability of the
autosampler and detector, accuracy of the data processing
unit and its processing parameters.
Summary of Contents for CHROMASTER 5160
Page 244: ......
Page 406: ......
Page 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...