3.4.1
Press [numerical] key (within [1] to [3]), and then the [ENT]
key.
During spectrum acquisition, the following screen is
presented.
SCANNING SPECTRUM NO.: 1 MODE: D2 REF
WL: *** nm
After spectrum acquisition, the monitor screen automatically
appears.
(d) Storing the Reference Energy with W Lamp in Memory (“4”
key followed by ENT key)
Set the minimum wavelength of scan range.
SCANNING WL RANGE
(WL1 WL2) nm W REF
<200 – 850>
2
00 – 600
Press [numerical] keys (within [2][0][0] to [8][5][0]), and then
the [ENT] key.
* Enterable range: 200 to 850 nm, enterable in 1-nm
steps
Set the maximum wavelength of scan range.
SCANNING WL RANGE
(WL1 WL2) nm W REF
<250 – 600>
200 – 6
00
Press [numerical] keys (within [2][5][0] to [9][0][0]), and then
the [ENT] key.
* Enterable range: Minimum scan range is 50 nm.
Measuring span is max. 400 nm.
MEMORIZE SPECTRUM NO. (1–3)
_
Input the number of spectral data.
Press [numerical] key (within [1] to [3]), and then the [ENT]
key.
3 - 33